GasAlertMicroClip XT
Maintenance
Note
Detectors that are configured for 1, 2, or 3 gases may
contain a dummy sensor in one of the four sensor locations.
3. Insert the new sensor(s). For sensor positioning, refer to
Figure 12.
Figure 12. Sensor Positioning
4. To complete the detector, refer to Reassembling the
detector.
Replacing the Oxygen Sensor
Note
Detectors that are configured for 1, 2, or 3 gases may contain a
dummy sensor in one of the four sensor locations.
1. Gently remove the circular rigidified flex PCB atop the
sensor from the metal sensor posts. Take care not to tear
the flex cable.
2. Note the placement of the PCB to ensure it is replaced
correctly. Remove the two screws on the PCB.
a Caution
Ensure no damage occurs to the battery.
3. Lift the PCB straight up. The oxygen sensor will stay in
the front shell. Remove the sensor.
4. On roughly the same spot on the front shell, place the
new sensor. Lower the PCB over the oxygen sensor.