GasAlertMicroClip XT
Technical Reference Guide
20
Figure 3. Connecting the IR Link
1. Activate the detector and wait for the startup tests to
complete.
2. Connect the USB cable to the USB port on the computer.
3. Connect the USB cable to the IR Link.
4. Insert the IR link onto the IR interface on the back of the
detector.
5. Open Fleet Manager II and access the Sensors tab.
Refer to Installing Fleet Manager II
.
6. From the Sensors tab, click Retrieve from Device at the
bottom of the window.
The fields will populate with the detectors current configurations.
7. Refer to the descriptions in the following sections to
define settings and enable/disable options:
Detector Identification
Sensor Configuration (CO, O
2
, H
2
S, and LEL)
User Options
Language Menu
8. When configuration of new settings is complete, click
Save to Device at the bottom of the window. The detec-
tor automatically updates with the new settings.