Visually inspect the battery to ensure no damage has
occurred.
When replacing the rear shell, ensure the charging pins
(bottom of inside rear shell) are aligned with the corre-
sponding holes on the PCB.
Press the front and rear shells together firmly to ensure
a proper seal. Ensure the front and rear shell have a
uniform, tight 1/16 in. (1 mm) seal on all sides of the
detector.
When replacing the screws, they must be seated prop-
erly to prevent cross threading. Turn the screw counter-
clockwise until a click is heard and then begin tightening
the screw clockwise.
Note
Ensure the rib on the interior rear shell inserts
between the battery and the PCB. Refer to the
following illustration.
New sensor(s) should be calibrated. Activate the detec-
tor and calibrate the sensor(s). Refer to Calibration
.