GasAlertMicroClip XT
Maintenance
15
To replace a sensor or sensor filter, refer to the following
illustration, table, and procedures.
1. Deactivate the detector. On a clean surface, place the
detector face down.
2. Remove the six machine screws from the rear shell.
Remove the back cover by lifting the top and the
bottom upwards simultaneously to prevent damaging
the charging pins.
Item Description
1 Front shell
2 LEL sensor
3PCB
4 PCB screws (2)
5 Rear shell
6 Machine screws (6)
7 Sealing rib
8 CO sensor
9H
2
S sensor
10 O
2
sensor
11 Sensor filter