GasAlertMicroClip
Operator’s Manual
14
Alarm Display Alarm Display
Sensor Alarm
During startup Error [sensor name]
displays
During normal operation Err displays
Confidence Beep and IntelliFlash
One beep and flash every second
Note: Confidence Beep and
IntelliFlash automatically
deactivate during low battery
alarm, self-test fail, calibration fail,
bump test fail, or an alarm event.
Low Battery Alarm
Sequence of 10 rapid sirens and
alternating flashes with 7 seconds of
silence in between (continues for
15 minutes)
and
L
flash,
LOW BAT
displays, and the vibrator alarm activates
After 15 minutes of the Low Battery
alarm, the Automatic Shutdown Alarm
sequence begins
OFF displays before deactivating
Automatic Shutdown Alarm
Sequence of 10 rapid sirens and
alternating flashes with 1 second of
silence in between (sequence reactivates
seven times)
LOW BAT and L display, and
vibrator alarm activates
OFF displays before deactivating
Note: Alarms can be set to latching or non-latching. To enable/disable
latching, select/deselect Latching Alarms in Fleet Manager II. Local
regulations may require a latching alarm.
If Low Alarm Acknowledge is enabled and a low alarm occurs, press
C to disable the audible alarm. The visual and vibrator alarms remain
activated. If the alarm escalates to a high, STEL, or TWA alarm, the
audible alarm reactivates.
Vue de la page 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21 ... 27 28

Commentaires sur ces manuels

Pas de commentaire

Edimax Technology PS-1206U manuály

Uživatelské manuály a uživatelské příručky pro Tiskárny Edimax Technology PS-1206U.
Poskytujeme 1 manuály pdf Edimax Technology PS-1206U ke stažení zdarma podle typů dokumentů: Uživatelský manuál


Tabulka s obsahem





Další produkty a příručky pro Tiskárny Edimax Technology

Modely Typ Dokumentu
Žádný