GasAlertMicroClip
User Manual
Note
The sensors are continuously tested while the
detector is activated.
Automatic Zero and O
2
Calibration (optional)
5. The sensors (H
2
S, CO, and LEL) are automatically
zeroed and the O
2
sensor is automatically
calibrated (unless it is disabled in the detector’s
options). The detector beeps twice to signal a
successful zero and span.
If only the automatic oxygen calibration is enabled on
the detector, the following screen appears.
Note
If ambient air is set to be measured 20.8% vol., the
automatic oxygen calibration screen displays 20.8
instead of 20.9.
Calibration Due Date (optional)
6. The LCD displays the number of days remaining
until the next calibration is due.
Note
If the
calibration interval is set to 0, then the
calibration due date start-up feature is bypassed.
If any sensor is past its calibration due date, the
detector displays CAL DUE todAY as the detector
beeps, flashes, and vibrates.
12
Vue de la page 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 56 57

Commentaires sur ces manuels

Pas de commentaire

Edimax Technology PS-1206U manuály

Uživatelské manuály a uživatelské příručky pro Tiskárny Edimax Technology PS-1206U.
Poskytujeme 1 manuály pdf Edimax Technology PS-1206U ke stažení zdarma podle typů dokumentů: Uživatelský manuál


Tabulka s obsahem





Další produkty a příručky pro Tiskárny Edimax Technology

Modely Typ Dokumentu
Žádný